OPR360US H0711
4-16
4
Setting Up For Patient Use
Event History Screen
Refer to Foldout F-9
From the Extended Functions menu touch the Event History button
to access the Event History Log. This log records the 1000 most
recent occurrences of alarm violations and settings, ventilator setting
changes, sensor calibration results and power On/Off sequences with
the date and time of each event. Recorded events are
color-coded - alarm violations and power OFF are red, setting
changes and calibration results are blue, and power ON and Start
Ventilating are in green. The Event History is retained after shutdown.
The software version (i.e. USR7.0) and the serial number (i.e. NXXX...)
are listed at the top of the Event History Log.
See “Save Feature” and “Download Feature” on following pages in
this section for details on saving Event History files.
Numeric Screen
Refer to Figure 4-6
Touching the Numeric button from the Main Screen menu displays
the following information:
Figure 4-6 Numeric Screen