Cleaning and Maintenance
6
and will depend on observance of safe handling precautions and the
ability to calibrate the sensor. Always make sure that the flow sensor
is completely dry before installation.
Exhalation Valve
The following diagram and instructions are for removal and
reinstallation of the exhalation valve.
1. Open the front panel door to expose the exhalation valve.
2. Remove the exhalation valve by releasing the latch at the top of
the panel and pulling the valve away from the ventilator. Refer to
Figure 6-5.
3. With a twisting motion, pull the plastic flow sensor away from the
outlet of the exhalation valve.
4. While holding exhalation valve body securely, turn the retaining
ring counter clockwise and remove it. See Figure 6-6.
5. Separate the exhalation valve cap from the valve body.
6. Remove diaphragm poppet assembly from the valve cap (do not
disassemble the diaphragm poppet assembly).
Figure 6-5 Remove Exhalation Valve