OPR360US H0711
5-5
Alarms
5
Alarm Tones
See Figure 5-4.
Touch the Alarm Tones button to open the screen. Use the Adjustment
knob to select one of three unique sets of sounds that will be activated
when an alarm is violated. Press Accept to confirm the change.
Exiting Alarm Screens
Touch any menu button on the control panel to exit any alarm screen.
See Figure 5-1
Front Panel Alarm Interface Environment
Alarm Silence Button
See Figure 5-1
Press the Alarm Silence button to mute silenceable alarms for two
(2) minutes and also to cancel the Shutdown Alarm after powering
down the ventilator. The indicator lamp on the Alarm Silence button is
lit when alarms are silenced. To cancel Alarm Silence press the Alarm
Silence button again.
Press the Alarm Silence for one (1) second or longer to enable the
Suction Disconnect feature.
NOTE: Pressing the Alarm Silence button does not mute a Device
Alert until the ventilator is powered down.
NOTE: The optional External Alarm Silence cable provides the same
function as the button except that it does not silence a the Shutdown
Alarm.
Figure 5-4 Alarm Tones