2-7
Overview
2
Patient and Breath Type/Mode Selection
At the far left of the Status Bar an icon is displayed that represents
which Patient Category and which mandatory Breath Type/Mode are
selected. See Section 4 for more details.
Patient Trigger Indicator
While ventilating, the Patient Category and mandatory Breath Type/
Mode selection area flashes green each time the patient triggers the
ventilator.
Int. Battery Charge Level
When the ventilator is powered by the internal battery an icon at the
far right of the Status Bar shows the remaining battery power. Each lit
bar represents approximately 25% of the total battery capacity.
Ext. Battery Indicator
When the ventilator is powered by an external battery “Ext Bat” is
displayed on the far right of the status bar.
Date/Time
Date and time are displayed in the far right corner of the display. The
date, time, and preferred format can be set in the Technical screen.
See Section 4 for more details.
Hour Meter
Touch the area just below the Date/Time to display the total
working hours of the ventilator. After 10 seconds the hour meter will
disappear.
Internal Battery
The e360 Ventilator is equipped with an internal battery that when
fully charged can support approximately one hour of ventilator
function at the following settings: Adult, SIMV with RR 15, VT 500,
t insp 1.0 seconds, FIO2 .30, PS 0, PEEP +5, Pause OFF, Sigh OFF,
Square wave form.
When the ventilator is operating on the internal battery:
o the Int Battery LED on the front panel lights and an audible alarm
sounds every 5 minutes
o the Battery Charge Level icon indicates the relative charge level of
the internal battery
The internal battery recharges whenever the ventilator is connected
to AC power, regardless of whether the power switch is ON or OFF.