OPR360US H0711
2-8
2
Overview
The internal battery requires up to 5 hours recharge from AC power
to obtain an 80% charge and is fully charged after 14-16 hours.
Recharge interval should not exceed 2 months.
If the internal battery voltage remains low (a Low Battery alarm
sounds after disconnecting from AC power) after a 5 hour charge on
AC power, the internal battery may need to be replaced. The standard
replacement schedule for the e360 internal battery is every 24
months. Refer to the e360 Ventilator Service Manual for replacement
procedure.