Alarms
Disconnect as well as all messages are displayed in the center
section of the status bar in the Graphical User Interface (GUI).
These messages flash while active then remain in a steady state
(latched) after the alarm violation has been corrected until the Reset
button is used to clear the alarm. Alarm messages are color-coded
and listed in order of priority. Red messages are alarms of high
priority such as Low MVE or High Paw. Messages displayed in yellow
are medium priority warnings such as Back-up ventilation. Green
Messages are low priority informational messages such as Ventilation
Suspended or Low Paw Below PEEP setting violation.
Device Alert LED
Refer to Figure 5-1.
Located on the left side of the control panel, the Device Alert LED
lights and a non-silenceable alarm sounds when there is a device
alert condition such as an empty battery or device malfunction. Refer
to table 5-2 for more information on device alert conditions.
1. Alarm Silence/Suction Disconnect button and indicator
2. Alarm Reset button
3. Mains Power, Internal Battery, and Device Alert indicators
4. Alarm Messages display area
5. Alarms Screen button
6. Touch buttons:
Alarm History button
Alarm Loudness button
Alarm Tones button
Save button
7. Setting adjustable alarms area
Figure 5-1 Alarms Screen Environment
1
2
3
4
5
6
7