OPR360US H0711
4-5
Setting Up For Patient Use
4
To Disable the O2 Sensor
1. Touch the Sensors button, then touch the O2 Sensor button.
2.
Touch Disable to disable the O2 sensor, then touch Yes to confirm.
3. Touch Exit.
To Enable the O
2 Sensor
1. Touch the Sensors button, then touch the O2 Sensor button.
2. Touch Enable to enable the O
2 sensor.
3. Touch Exit.
Patient Setup
Refer to Foldout F-11.
Press the Setup & Calibration menu button, then touch the Patient
Setup button to access this screen.
Patient Category
Choose between Adult and Ped/Infant patient category. The setting
impacts ventilation and alarm settings ranges and ventilation
management algorithms. If any alarm or ventilator setting is out of
range after changing the Patient Category, the Alarms and Messages
display shows “[Setting] Out of Range” and the LED display for the
parameter(s) that are out of range flashes and an audible alarm sounds.
NOTE: Always select Ped/Infant category when using a pediatric or
infant breathing circuit.
Weight Units
Select either Lb or kg for the unit of measure for weight.
Ideal Weight
Enter the ideal patient weight value, between 1 – 999 kg/ 2 - 2202 lb.
The ideal weight must be entered before you can choose to display
exhaled volume measurements in mL/lb or mL/kg.
Volume Units
Choose mL or mL/kg for the unit of measure for the VT
E display. You
must enter an ideal weight before selecting mL/Kg.
NOTE: Selecting the volume unit affects the numeric data display for
VT
E only.
Figure 4-2. Basic Data Set Bar indicating that the O2 sensor
and FIO2 alarms are disabled