Explanation of Modes, Breath Types and Special Functions
Explanation of Modes, Breath Types and Special Functions
7
flow/volume that is added in and delivered to the patient in order to
compensateforthevolumethatis“lost”inthetubingissubtracted
from both the displayed values.
Caution The e360 Circuit Check determines how much the e360
compensates for circuit compliance. If you have a bigger circuit
and humidifier in place when you do the test it will result in more
adjustment. A smaller circuit/humidifier will result in less adjustment.
Make certain that the ventilator is set up exactly like it will be on the
patient (including filters and water in the humidifier) when you do
these tests or the delivered and displayed volumes will be incorrect
during ventilation.
Non Invasive Ventilation (NIV)
The e360 Ventilator can be used for invasive (intubated patient) or
noninvasive (mask) ventilation. When the Non Invasive button is
activated on the front panel (LED lights), e360 tailors the ventilator’s
performance (described below) to meet the needs of a patient who is
breathing from a mask rather than an invasive artificial airway or an
uncuffed/deflated cuff trach tube. Non Invasive can be used with any
mode of ventilation. It is factory preset to OFF and the setting returns
to OFF after power down.
Leak Compensation in NIV
The Non Invasive function automatically provides leak compensation/
baselinepressuremanagementwithabiasowrangeof3to25L/min
in order to accommodate the potential for bigger airway leaks around
the non-vented mask or uncuffed/deflated cuff trach tube. (When Non
Invasive is not activated and Leak Comp is ONbiasowisonly3-8L/
minPed/Infantand3-15L/minAdult.)
Alarm Settings in NIV
The Low MVE and the Disconnect Threshold alarms, can be set to
OFF while Non Invasive is activated. All other alarms are functional.
If the Low MVE or Disconnect Threshold alarm is OFF when Non
Invasive is deactivated, the alarms are automatically turned back
on and the Low MVE alarm is set to the lowest value while the
Disconnect Threshold alarm is set to the highest value.
NOTE: To minimize the chances of auto-triggering due to leaks,
Newport Medical recommends using Pressure trigger (start at 2
cmH2O/mbar for Adult and 1 cmH2O/mbar for Ped/Infant) when
using the e360 for noninvasive mask ventilation. Use a nonvented
mask to ensure proper patient-ventilator synchrony.