OPR360US H0711
8-17
Specifications
8
Physical Specifications
Table 8-3
Description
Power
Specifications
AC input range: 100 to 240 VAC, 250 VA maximum,
50/60 Hz (±10%), 2A for 125 VAC, 1A for 250 VAC
Internal battery: Fully charged battery can support
approximately one (1) hour of complete ventilator
function at the following standard settings: Adult, VC/
SIMV, VT 500, FiO
2 .30, Insp Time 1.0 s, Resp Rate 15,
PS 0, PEEP +5,
Pause Off, Sigh Off, Square Wave.
The ventilator recharges the internal battery whenever AC
power is connected (whether the power switch is ON or
OFF).
Minimum recharge time:
From Low Battery Alarm to Full = 5 hours
From Empty to Full = 14 - 16 hours
Power Cord Requirements:
In the USA, power cord must comply with UL2601.
For 125 VAC, 15 A: 2-pole, 3-wire, 18 AWG, grounding-
type, 5-15P hospital-grade plug cap, < 10 ft (3 m) long,
CSA and UL-approved or for 250 VAC, 15 A: 2-pole,
3-wire, 18 AWG, grounding-type, 6-15P, hospital-grade
plug cap, < 10 ft (3 m) long, CSA and UL-approved
Compliance
Complies with IEC 60601-1 with amendments 1 & 2,
C22.2 No 601.1-M90
and UL Std No 2601-1
Type B Applied Parts
Rated battery voltage: 12 VDC
Class I electrical-safety equipment
Mode of operation: continuous operation (as per Clause
5.6)
IPX0 rating (degree of protection against ingress of water)
Expiratory
Channel
Resistance
Pressure drop less than 1.7 cmH
2O/mbar @ 50 L/min
Adult
less than 1.7 cmH2O/mbar @ 20 L/min Infant
NOTE: Testing was performed according to ASTM
F1100-90
Dimensions
Height: 14 in (35.56 cm)
Width: 12 in (30.48 cm)
Depth: 14 in (35.56 cm)
Weight: 40 lbs. (18.14 kg)
Display
6.4 in active matrix color LCD
Touch screen transparent plastic-covered glass.
Pressure-sensitive surface can electronically decode
touch position.