OPR360U A0509
5-1
5
OPR360U A0509
5-1
5
OPR360U A0509
5-1
5
OPR360U A0509
5-1
5
OPR360US H0711
2-3
Overview
2
Touch a parameter control on the GUI or press a parameter
membrane button on the Control Panel; then,
Turn the Adjustment knob (Foldout F-4, Item 9) to make a change;
and,
Press the Accept button (Foldout F-4, Item 10) to confirm and invoke
the change.
If the Accept button is not pressed within 10 seconds, the setting will
not be changed and will revert to the previous condition/value.
Selecting Mandatory Breath Type and Mode
Refer to Foldout F-4, item 12.
NOTE: More specific information on modes and breath types is
available in Section 7.
The mandatory breath type and mode are displayed in the upper left
corner of the GUI. Spontaneous breath type is pre-selected based
on the mandatory breath type.
Modes are selected by pressing the Volume Control or Pressure
Control breath type buttons repeatedly until the desired mode is
highlighted.
Biphasic Pressure Release Ventilation (BPRV) is selected by choosing
Pressure Control A/CMV or SIMV mode and then selecting Open
Exhalation Valve from the Advanced Data Set GUI screen. See Data
Sets, Section 4, for more information.
Volume Target Pressure Control (VTPC) is selected by choosing a
Volume Control or Pressure Control A/CMV or SIMV mode and then
selecting Volume Target ON from the Advanced Data Set GUI screen.
See Advanced Data Sets in Section 4 for more information.
See the Ventilation Controls Guide in Section 4 for a full list of breath
type and mode combinations.
Non Invasive (Ventilation) (NIV) Function
Refer to Foldout F-4, item 11.
To activate the NIV function in any mode or breath type, press the
Non Invasive button (the indicator will light) and then press Accept.
Non Invasive always reverts to OFF when e360 is powered down (the
setting is not retained). See Section 7 for more information on Non
Invasive function.