OPR360US H0711
2-4
2
Overview
Patient Triggering Methods
Refer to Foldout F-4, item 8.
The ventilator offers the clinician the choice of flow or pressure (P)
triggering for patient-initiated breaths in all modes of ventilation. To
select Flow or P, press the Trig button (LED lights) and then press
Accept.
Basic Ventilation Controls
Refer to Foldout F-4, item 8.
To set FIO2, Tidal Volume, Flow, Inspiratory Time (t Insp), Respiratory
Rate (Resp Rate), Pressure Support, Pressure Limit, PEEP/CPAP and
Trigger (flow or pressure (P)):
1. Press the button below the corresponding display to select a
parameter.
2. Rotate the Adjustment knob to adjust the setting while it is
flashing.
3. Press the Accept button to invoke the change.
Or
Select and adjust multiple basic ventilation controls within 10
seconds of the last change and then press the Accept button to
invoke all of the changes. The display(s) will stop flashing and the
setting(s) will take effect.
Flow or Inspiratory Time in Volume Control
When ventilating in Volume Control, you can choose to set either
Flow or Inspiratory Time (t Insp) for mandatory breaths. For all other
mandatory breath types, only t Insp can be adjusted. Press the Select
button to toggle between Flow and t insp.
Manual Inflation Button
Refer to Foldout F-4, item 11.
Press and hold the Manual Inflation button to deliver a manual
inspiration. The inflation ends when the user releases the button, five
seconds elapses, or a High Paw alarm is violated, whichever occurs
first.