EasyManua.ls Logo

Newport e360 - Alarms, Controls, Monitored Data, Setup & Calibration

Newport e360
150 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
OPR360U A0509
5-1
5
OPR360U A0509
5-1
5
OPR360US H0711
8-1
Specifications
8
Alarms, Controls, Monitored Data, Setup & Calibration
Table 8-1
Item
Location and/or
Function
Range and Resolution or
Description
All Other
Messages
GUI-
Informational
Message
For any message not shown here
refer to Section 5, Alarms for more
information.
Alarm History
GUI-Alarms See Event History
Alarm
Loudness
GUI-Alarms
Range: 1 to 10 (55 - 75 dbA)
Default : 5
Accuracy: +/- 5 dbA
(Alarm) Reset
Control Panel
Clears visual indicators and messages
for alarms that are no longer violated.
Alarm Silence
Control Panel
Mutes silenceable, audible alarms for
two (2) minutes. Pressing the Alarm
Silence button again turns off the
alarm silence function. This button
will not silence a Device Alert alarm
without first powering the ventilator
OFF.
Alarm Tones
GUI-Alarms
Range: 1-3
Default: 1
Apnea
GUI-Alarms-
Adjustable
Alarm is violated when no breath or
effort is detected within the set apnea
interval.
Range: 5 to 60 seconds
Accuracy: ±1 second
Back Up
Ventilation
(BUV)
(Results from
low MVE Alarm
Violation)
Alarms-Non-
Adjustable
If the current mode is A/CMV or SIMV,
Back Up Ventilation employs the
current Control Panel settings except
for Respiratory Rate which increases
to 1.5 times the current setting (15 b/
min minimum, 100 b/min maximum).
If the current mode is SPONT, the
ventilator delivers pressure control
mandatory breaths with the following
settings:
• Plimit 15 cmH
2O/mbar above PEEP
setting
• t Insp 0.6 seconds Ped/Infant, 1.0
seconds Adult
• Resp Rate 20 b/min Ped/Infant,
12 b/min Adult
(Cont. next page)