Setting Up For Patient Use
4
Technical
The Technical Screen is accessed from the Setup & Calibration
screen. Refer to Foldout F-10 for a sample of the Technical screen.
Set ventilator specific technical settings that are appropriate for your
hospital or patient.
Comm (Communication) Protocol
Select the RS232 communication protocol that corresponds with the
monitoring system that is connected to the e360 Ventilator. Touch
the button to select from Newport, Newport 2, and Vuelink.
Contact Newport Technical Service for more details regarding the
communication port protocol.
Display Brightness
Adjust the built-in GUI display brightness.
Regional Settings
• Altitude
Adjust the altitude setting in 200-meter increments so that it
corresponds with your local ambient altitude. The altitude setting may
be adjusted up to 4,000 meters (13,124 feet).
Date Format
Select from three (3) date formats: month–day–year, day-month-year
or year-month-day.
Date and Time
Set the Month, Day, Year and Time.
Screen Files
Touch this button to open the Screen Files List window which
contains the last 200 saved screen images (.bmp files). Ensure that a
flash drive is properly inserted into the USB port. Use the Adjustment
knob to scroll to a .bmp file, then touch Download button to download
the file to the flash drive. A File Storing Successful message will
appear with a short two-toned beep when the download to the flash
drive is complete. Repeat the process for each file to be downloaded.
Event History Files
Touch this button to open the Event History Files List window which
contains the last 200 Event and Alarm History Logs (.csv files).
Ensure that a flash drive is properly inserted into the USB port. Use
the Adjustment knob to scroll to a file, and then touch Download
button to save the file to the flash drive. A File Storing Successful
message will appear with a short two-toned beep when the download
to the flash drive is complete. Repeat the process for each file to be
downloaded.