OPR360US H0711
4-8
4
Setting Up For Patient Use
NOTE: See “Save Feature” and “Download Feature” on following pages
in this section for details on saving Screen and Event History files.
Ventilation Controls Guide
The following table shows the ventilation parameters that are active
in each mode/breath type.
X = active, D =dimmed (not active but still adjustable)
Control
Display
VC/
ACMV
VC/
SIMV
VC/
SPONT
PC/
ACMV
PC/
SIMV
PC/
SPONT
F
IO2
X X X X X X
Tidal
Volume
X X D D D D
Flow
X X D D D D
t Insp
X X D X X D
Resp. Rate
X X D X X D
Pressure
Support
D X X D X X
Pressure
Limit
D D D X X D
PEEP/CPAP
X X X X X X
Trig
X X X X X X
Slope/Rise
D X X X X X
Exp. Thresh
D X X D X X
Pause
X X D D D D
Open Exh
(ON/OFF)
D D D
X
ON =
BPRV
X
ON =
BPRV
D
Flow Wave
X X D D D D
Volume
Target (ON /
OFF)
X
ON =
VTPC
X
ON =
VTPC
X
ON =
VTPC
X
ON =
VTPC
X
ON =
VTPC
X
ON =
VTPC
Mode LEDs
Volume
Control
and
ACMV
Volume
Control
and
SIMV
Volume
Control
and
SPONT
Pressure
Control
and
ACMV
Pressure
Control
and
SIMV
Pressure
Control
and
SPONT
Table 4-1: Ventilation Controls Guide