OPR360US H0711
2-5
Overview
2
O2 (3 min) Button
Refer to Foldout F-4, item 11.
Press the O2 (3 min) button to start a timed delivery of 100% oxygen,
regardless of the current FIO2 setting. The indicator on the O2 (3
min) button lights when this function is activated. F
IO2 returns to the
set value and the indicator turns off after three minutes or when the
button is pushed a second time, whichever comes first.
Alarm Silence Button
Refer to Foldout F-4, item 1.
Press the Alarm Silence button (LED lights) to mute silenceable
audible alarms for two (2) minutes or to cancel the Shutdown alarm
after the power is switched to OFF. Press again to cancel alarm
silence.
Suction Disconnect Function
Prior to a planned circuit disconnect, press and hold Alarm Silence
button for one second until a second tone sounds to activate the
Suction Disconnect function. See Section 5, Alarms, for more details.
Alarm Reset
Refer to Foldout F-4, item 2.
Press the Reset button to clear visual indicators for alarms that are
no longer violated.
Graphical User Interface (GUI) Screens
Refer to rear Foldout F-6 - F-8 for samples of these screens.
Alarms Screen
Refer to Foldout F-4, item 5.
Press the Alarms Screen menu button on the Control Panel to open
the Alarm Settings screen on the GUI. From this screen the user
can modify all adjustable alarm settings, view Alarm History, adjust
Alarm Loudness and Alarm Tones and Save the screen image for later
downloading. See Section 5, Alarms, for more details.