OPR360U A0509
5-1
5
OPR360U A0509
5-1
5
OPR360U A0509
5-1
5
OPR360U A0509
5-1
5
OPR360US H0711
6-7
Cleaning and Maintenance
6
Inspiratory Manifold
The following diagram and instructions are for removal and
reinstallation of the inspiratory manifold.
1. Using a flat screwdriver or a coin, unscrew the two screws on
the lower right front panel section and remove it to expose the
inspiratory manifold. See Figure 6-8.
2. Use a Philips screwdriver to remove the upper and lower manifold
retaining screws and pull the manifold away from the ventilator.
3. Remove the manifold cap by removing the four screws to expose
the diaphragm. Handle with care. See Figure 6-9.
4. Remove the inspiratory port from the block by turning it
counterclockwise.
5. To reassemble and reinstall, reverse the above steps.
Figure 6-8 Remove Panel Section
Figure 6-9 Inspiratory Manifold Block