Unpacking, Assembly, and Safety Check
3
8. PresstheReset button to clear the High Paw alarm message and
visual indicators.
Minute Volume / Back Up Ventilation / Apnea Alarms
1. AdjusttheResp Rateto20b/min.Verifythatbothaudibleand
visual indicators for High MVE(ExhaledMinuteVolume)alarms
are activated within 30 seconds.
2. AdjusttheResp Rate backto10b/min.Verifythatwithin30
seconds the audible alarm stops and the visual alarm indicator
is steadily lit. Press the Reset button to clear the High MVE
alarm message and indicator.
3. AdjusttheResp Rateto1b/min.Verifythatwithin30seconds
the audible and visual alarm indicators for APNEA and the
Low MVEalarmareactivated.Verifythatafter65seconds,
Back Up VentilationbeginsandisindicatedwithaBackUp
VentilationmessageontheAlarmsandMessagesdisplay.
4. Verifythatwithin30secondsfollowingthestartofBackUp
Ventilationthealarmindicatorsaresteadilylitsignifyingtheend
ofBackUpVentilation.AdjusttheResp Rateto10b/min.
Press the Reset button to clear the alarm messages and
indicators.
Trigger/Pressure Support
1. Set High MVEalarmto12L,ModetoSPONT, Pressure Support
to 10 cmH2O/mbar,PEEP to 3 cmH2O/mbar,P Trigto2.0
cmH2O/mbar.
2. Brieysqueezethetestlungtocreateanegativepressureinthe
breathingcircuit.Verifythatthegreenpatienteffortindicator
“blinks”,andthatapressuresupportbreathisdelivered.
3. Select Flow Trigandsetto2.0L/minandrepeatstep2.
4.
Set Mode to A/CMV, PEEP to 0 cmH2O/mbarandTrigtoP=5.0
cmH2O/mbar.Allothercontrolsshouldstillbeatstandardsettings.
Volume/Flow/Rate Accuracy Test
1. Change the GUI to the Numeric screen(viaMainScreenbutton)
toseethemonitoredexhaledtidalvolume(VTE).Verifythatthe
monitoredvalueiswithin±20%ofthe Tidal Volume setting on
the front panel.
2. ObservetheInspFlowdisplayontheNumeric screen and verify
thatthemeasuredvalueiswithin±5L/minoftheFlow setting on
the front panel.
3. ObserveRRtot(monitoredtotalbreathrate)ontheNumeric
screen and verify that after 30 seconds the monitored value is
within±1b/minoftheRespRatesettingonthefrontpanel.