Setting Up For Patient Use
4
Power Conditions
Off: Power switch is in the OFF position.
Ventilation Standby: Power switch is in the ON position but the
ventilator is not yet in the ventilating condition.
Ventilating: Power switch is in the ON position and the user
has touched the Start Ventilating button.
WARNING! Never connect the patient breathing circuit to the patient
while the ventilator is in ventilation standby condition. Always touch
the “Start Ventilating” button on the GUI prior to connecting the
breathing circuit to the patient.
Shutdown Alarm
After powering off the ventilator it will initiate a Shutdown alarm as
confirmation that the ventilator has been powered OFF. Cancel the
Shutdown alarm by pressing the Alarm Silence button.
NOTE: Make sure that the e360 is completely shutdown before the
power switch is turned on again. If the power switch is turned off and
then on too quickly and the screen does not load correctly, turn the
power off and wait 15 seconds before powering ON the ventilator.
Overview: Preparing For Patient Ventilation
1. Attach the breathing circuit and humidifier. Fill the humidifier
water chamber.
2. Connect the air and oxygen hoses to the appropriate source gas
supplies.
3. Plug the ventilator power cord into AC power.
4. Turn the ventilator ON. It will be in the standby condition.
5. While in standby:
a. Perform a Circuit Check (follow the on-screen instructions).
b. Touch Sensors and perform the O
2 Sensor calibration.
WARNING! Newport Medical recommends that you perform the
Circuit Check (which includes an Exhalation Flow Sensor Calibration)
and O
2 Sensor Calibration procedures before connecting the
ventilator to a patient.
c. Touch Patient Setup and enter selections for:
Patient Category
Ideal Weight and Units of Measure
Circuit Type
Leak Compensation (Newport recommends that you always
keep Leak Compensation turned ON)
Compliance Compensation (ON or OFF)
Sigh (ON or OFF)
d. Verify that ventilation parameters settings, including
Advanced settings are appropriate.