3
Unpacking, Assembly, and Safety Check
3. AdjusttheFIO2to.23.Disconnectthehighpressureoxygenhose
fromthegassource.Verifythate360providesanaudiblealarm
and visual O2 Supply Loss alarm message.
4. Reconnect the high pressure oxygen hose to the gas source.
Verifythatthealarmisnolongerviolated.PushReset to clear
visualmessageandindicator.ReturntheFIO2to.21.
5. Disconnectthehighpressureairhosefromthegassource.Verify
that e360 provides an audible alarm and visual Air Supply Loss
alarm message.
6. Reconnectthehighpressureairhosetothegassource.Verify
that the alarm is no longer violated. Push Reset to clear visual
message and indicator.
AC Power Loss/Battery Backup Alarm
1.
Whiletheventilatorisoperating,unplugtheACpowercord
fromthewalloutlet.Verifythattheventilatorcontinuesfunctioning
andprovidesanaudibleandavisualalarm,theInt Battery
(InternalBattery)indicatorlightsandthemessageAC Power Loss
Battery Back Up appearsinthewindow.Thee360issuesashort
beepeveryve(5)minuteswhilerunningoninternalbattery.
2.
PlugtheACpowercordbackintothewalloutlet.Verifythatthe
ventilatorcontinuesfunctioning,theInternalBatteryindicatorgoes
out and the Mains(BatteryCharging)indicatorlights.
3. Push Reset to clear visual messages and indicator.
High/ Low Airway Pressure Alarms / Circuit Disconnect Alarm/
Alarm Silence
1. Removethetestlung.Verifythatbothvisualandaudible
indicators for the Low Paw(lowairwaypressure)alarmare
activated after two mandatory breaths and after three breaths the
Circuit Disconnect alarm message is displayed.
2. PresstheAlarm Silence button and verify that the audible alarm
is muted but the alarm lamp continues to blink and the alarm
message remains displayed.
3. Press the Alarm Silence button again and verify that the audible
alarm resumes beeping.
4. Re-attachthetestlung.Verifythattheaudiblealarmstopsand
the360°alarmlampissteadilylit(latched).
5. Press the Reset buttontoclearallvisualalarmindicators(lamp
andmessages).
6. Remove the test lung and occlude the patient wye connector
ofthebreathingcircuit.Verifythatboththevisualandaudible
indicatorsforHighPaw(highairwaypressure)alarmare
activated.
7. Re-attachthetestlungtothewyeconnector.Verifythatthe
audible alarm stops and the alarm indicator is steadily lit.