Unpacking, Assembly, and Safety Check
3. AdjusttheFIO2to.23.Disconnectthehighpressureoxygenhose
fromthegassource.Verifythate360providesanaudiblealarm
and visual O2 Supply Loss alarm message.
4. Reconnect the high pressure oxygen hose to the gas source.
Verifythatthealarmisnolongerviolated.PushReset to clear
visualmessageandindicator.ReturntheFIO2to.21.
5. Disconnectthehighpressureairhosefromthegassource.Verify
that e360 provides an audible alarm and visual Air Supply Loss
alarm message.
6. Reconnectthehighpressureairhosetothegassource.Verify
that the alarm is no longer violated. Push Reset to clear visual
message and indicator.
AC Power Loss/Battery Backup Alarm
1.
Whiletheventilatorisoperating,unplugtheACpowercord
fromthewalloutlet.Verifythattheventilatorcontinuesfunctioning
andprovidesanaudibleandavisualalarm,theInt Battery
(InternalBattery)indicatorlightsandthemessageAC Power Loss
Battery Back Up appearsinthewindow.Thee360issuesashort
beepeveryve(5)minuteswhilerunningoninternalbattery.
2.
PlugtheACpowercordbackintothewalloutlet.Verifythatthe
ventilatorcontinuesfunctioning,theInternalBatteryindicatorgoes
out and the Mains(BatteryCharging)indicatorlights.
3. Push Reset to clear visual messages and indicator.
High/ Low Airway Pressure Alarms / Circuit Disconnect Alarm/
Alarm Silence
1. Removethetestlung.Verifythatbothvisualandaudible
indicators for the Low Paw(lowairwaypressure)alarmare
activated after two mandatory breaths and after three breaths the
Circuit Disconnect alarm message is displayed.
2. PresstheAlarm Silence button and verify that the audible alarm
is muted but the alarm lamp continues to blink and the alarm
message remains displayed.
3. Press the Alarm Silence button again and verify that the audible
alarm resumes beeping.
4. Re-attachthetestlung.Verifythattheaudiblealarmstopsand
the360°alarmlampissteadilylit(latched).
5. Press the Reset buttontoclearallvisualalarmindicators(lamp
andmessages).
6. Remove the test lung and occlude the patient wye connector
ofthebreathingcircuit.Verifythatboththevisualandaudible
indicatorsforHighPaw(highairwaypressure)alarmare
activated.
7. Re-attachthetestlungtothewyeconnector.Verifythatthe
audible alarm stops and the alarm indicator is steadily lit.