OPR360US H0711
6-8
6
Cleaning and Maintenance
Oxygen Sensor
The following diagram and instructions are for the removal and
reinstallation of the oxygen sensor.
1. Set the power switch to the OFF position and disconnect the
ventilator from AC power and gas supplies.
2.
Using a screwdriver or a coin, unscrew the two screws on the
lower right front panel section and remove it to expose the oxygen
sensor. See Figure 6-11.
3. Locate the sensor cable and turn the twist collar counter
clockwise to remove the cable from the sensor.
4. Pull the sensor straight out to remove it. Discard oxygen sensor in
accordance with local regulations.
NOTE: The Inspiratory Manifold may need to be removed to aid in
sensor removal.
5. To install a new sensor, reverse the steps.
6. Calibrate the sensor as described in section 4.
NOTE: Before calibrating a new oxygen sensor, open the package
and allow the sensor to be exposed to room air for 20 minutes.
Fuses
The following diagram and instructions are for the removal and
reinstallation of fuses.
The fuses are located in the top of the AC power module on the rear
of the e360.
1. Set the power switch to the OFF position and disconnect the
ventilator from AC power and gas supplies.
2. Using a small flathead screwdriver, squeeze the tabs on the fuse
drawer to loosen it; then pull it from the AC power module. See
Figure 6-10.
3. Inspect and replace the fuses only if they are blown.
Figure 6-11 Remove Panel Section