EasyManua.ls Logo

Newport e360 - Emergency Intake Valve; Safety Check Procedure; Setup and Inspection

Newport e360
150 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
OPR360U A0509
5-1
5
OPR360U A0509
5-1
5
OPR360U A0509
5-1
5
OPR360U A0509
5-1
5
OPR360US H0711
3-8
3
Unpacking, Assembly, and Safety Check
WARNING! Use water traps or heated wires in appropriate locations
in the breathing circuit to prevent water from pooling in the circuit
tubing,drainingintothepatientairwayordrainingintotheventilator.
Emptyandcleanwatertrapsasnecessary.Neverdrainthewater
backintothehumidierwaterchamber.
CautionTheventilatorisreadyforoperationonlywhenitis
completely assembled and has successfully completed the Safety
Check and Circuit Check procedures.
Safety Check Procedure
Thee360doesaself-diagnosticstestwhenpoweredON which
veriestheoperationofinternalelectronics.NewportMedical
recommends that you perform a complete Safety Check prior to
the initial use of the ventilator and at least with every preventative
maintenance interval. Use the e360 Safety Check Record at the end
of this section to record the results of each check.
Donotusethee360ventilatorifitdoesnotpasstheSafetyCheck
Procedure.
Setup and Inspection
1. Assembletheventilatorsystem.
2. InspecttheNewporte360ventilator,Newportaircompressor(if
used),ACpowercords,andverifythatthereisnoevidenceof
wear or damage which might contribute to a malfunction.
3. ConnectACpowercord(s)toproperlygroundedwallreceptacles.
4.
Inspect the high pressure air and oxygen inlet water traps on the
back of the e360 to ensure that there is no water or debris present.
5. Ensurethatthehighpressureairandoxygenhosesarermly
securedontotheventilatorinletttings.
6. Attacharecommendedtwo-limb22mmbreathingcircuitand
lters.Havea500mL(or1Lenclosed)testlungavailable.
7. Inspect the patient breathing circuit and all connections to verify
that there is no evidence of wear or damage which might result in
leaksand/orcontributetoventilatormalfunction.
Emergency Intake Valve
1. Make sure the e360 power is turned OFF on the back of the
ventilator.
2. Verifythataircanbedrawnintothepatientbreathingcircuit
through the emergency intake valve. You may create a negative
effortontheinspiratorylimbofthepatientbreathingcircuitby(1)
usinga“bellows”typetestlung,or(2)inspiringthroughabarrier
lterontheinspiratorylimbofthepatientcircuit.

Table of Contents

Other manuals for Newport e360

Related product manuals