OPR360U A0509
5-1
5
OPR360U A0509
5-1
5
OPR360U A0509
5-1
5
OPR360U A0509
5-1
5
OPR360US H0711
8-2
8
Specifications
Item
Location and/or
Function
Range and Resolution or
Description
Back Up
Ventilation
(BUV)
(Results from
low MVE Alarm
Violation)
(cont.)
Changing any ventilation setting that
affects mode, breath timing,
flow/volume, pressure, or trigger
sensitivity suspends back up
ventilation for 60 seconds. It is also
suspended for 60 seconds following
Suction Disconnect function with
breathing circuit disconnect then
reconnect.
Back Up Ventilation terminates when
MVE exceeds the Low MVE Alarm
setting by 10%
Breath Type/
Mode
Setting
Monitored
Control Panel/
GUI
GUI-Monitored
Data
Breath Type:
Volume Control, Pressure Control or
Volume Target Pressure Control and
Biphasic Pressure Release Ventilation
(BPRV)
Mode:
A/CMV (assist/control mandatory
ventilation),
SIMV (synchronized intermittent
mandatory ventilation), or
SPONT (spontaneous ventilation)
Selected breath type and mode is
displayed below the Patient Category
icon on the upper left side of GUI.
Cdyn effective
(mL/ cmH
2O/
mbar) (Effective
dynamic
compliance)
GUI-Monitored
Data
Displayed range:
0 to 999.9 mL/ cmH2O/mbar
Accuracy:
± 1 mL/ cmH2O/mbar
Cdyn effective = VTE/ (Ppeak – Pbase)
NOTE: Calculated for time triggered
breaths only. VTE, Ppeak, and Pbase
must be valid to calculate Cdyn
effective.
Circuit
Disconnect
Alarms- Non
adjustable
Indicates that the Disconnect
Threshold % is greater than or equal
to the set value.
Circuit Type
GUI-Patient
Setup
Range: Heated Expiratory Limb,
Heated Inspiratory Limb, Heat
Moisture Exchanger (HME), and Test
Lung.
(Cont. next page)