OPR360US H0711
8-3
Specifications
8
Item
Location and/or
Function
Range and Resolution or
Description
Circuit Type
(cont.)
GUI-Patient
Setup
1. Heated Exp. Limb = heated
humidifier with dual heated wire
breathing circuit.
2. Heated Insp. Limb = heated
humidifier with no heated wires in the
circuit or heated wire in the inspiratory
limb of the breathing circuit only.
3. HME = no heated humidifier,
unheated breathing circuit with
passive humidification (heat moisture
exchanger [HME] or hydroscopic
condensing humidifier [HCH]).
4. Test Lung = no humidification, no
heat.
NOTE: Circuit Type selection affects
the accuracy of monitored values
for expiratory flow, expiratory tidal
volume and expiratory minute volume.
Comm.
Protocol
GUI-Technical
Range: Newport, Newport 2, Vuelink
Allows user to select protocol to
communicate with patient monitoring
systems via the RS232 port.
Compl Comp
(Compliance
Compensation)
GUI-Patient
Sertup
Range: ON or OFF
Volume Control mandatory
breaths: flow delivery is adjusted
to compensate for the effects of
breathing circuit compliance and
the display of monitored flows and
volumes approximate effective
delivered flows and volumes.
Cstat
(mL/cmH
2O/
mbar)
(Static
compliance)
GUI-Monitored
Data
Displayed range: 0 to 999.9 mL/
cmH2O/mbar
Accuracy: ± 1 mL/ cmH2O/mbar
Cstat = VTE/ (Pplat – PEEP)
Maneuver based with time stamp for
up to 24 hrs.
NOTE: Calculated for time-triggered
breaths only.
Cstat is updated immediately
following an update of Pplat (Insp
Hold Maneuver). It is also updated
following an update of Total-PEEP (Exp
Hold Maneuver) if this maneuver is
performed within 5 min. following Pplat
update. In this case, the Auto-PEEP
measurement is substituted for the
PEEP measurement in the equation.