OPR360US H0711
8-4
8
Specifications
Item
Location and/or
Function
Range and Resolution or
Description
Data Read
Failure Contact
Service Alarm
Alarms-Non
Adjustable
Software type is read invalid from both
the EEPROM on the main board and a
stored file on the compact flash card.
Device Alert
Control Panel-
Alarms-Non
Adjustable
LED on Control Panel lights (if possible).
Unsilenceable audible alarm sounds
if there is a ventilator malfunction
(messages associated with a device
alert are described in Section 5). Also
activated if less than 10% of internal
battery operation time remains.
WARNING! If a Device Alert occurs,
ventilation ceases and the emergency
intake valve and exhalation valve
opens.
Disconnect
Threshold
(VTE Var %)
GUI-Alarms-
Adjustable
Percent of difference between
inspiratory and expiratory tidal
volumes.
Range: 20 to 95%
Accuracy: +/- 10%
NOTE: When Non Invasive is activated
this alarm may be set to OFF.
Display
Brightness
GUI-Technical
Range: 0-100%
Default: 50%
Allows user to adjust display
brightness of GUI.
EEPROM
Read Error
Alarms-Non
Adjustable
Software type is read invalid from
the EEPROM on the main board. It
is logged in the event history page,
but no audible alarm and alarm LED
indicator will be activated. No error
message will be displayed on the
alarm/message bar on the screen.
Event History
(Alarm History)
GUI-Extended
Functions (and
Alarms)
A log of the past 1000 events. Events
are color coded:
Green: Power ON and Start
Ventilating.
Blue: Control settings and changes,
calibration results and Circuit Checks.
Red: Alarm violations and messages
and Power OFF
Event History is retained after Power
shutdown. Save Event History button
allows user to capture current event
log for later downloading in .csv
format.