OPR360US H0711
8-5
Specifications
8
Item
Location and/or
Function
Range and Resolution or
Description
Event History
Files
GUI-Technical
Contains a list of the last 200 saved
Event and Alarm History logs (.csv
files). Files are assigned an 8-digit file
name that includes a letter for file type
(H for history, etc), the last 4 digits
of the serial number, and a 3 digit
sequential number.
Expiratory
Hold
GUI-Extended
Functions
Range: Up to 20 seconds
Generates an Expiratory Hold
maneuver at the end of a mandatory
breath exhalation for as long as the
button is pushed.
Expiratory
Threshold
GUI-Advanced
Data Set
Range: AUTO 5 – 55%, (resolution
1%) of peak flow.
Manually or automatically sets the
flow cycling-off threshold for Pressure
Support and Volume Target Pressure
Support breaths
F
IO2 (Set)
Control Panel
Range: 0.21 to 1.00 (resolution 0.01)
Accuracy: ±.03
F
IO2
(Monitored)
GUI-Monitored
Data
Displayed range: 0.21 to 1.00
(resolution 0.01)
Accuracy: ±0.03 (“—-” displayed if
sensor is disconnected)
FIO2 High
Alarms-Non
Adjustable
Monitored FIO2 is more than 0.07
above the set FIO2 for 30 seconds.
Accuracy: ±0.03
F
IO2 Low
Alarms-Non
Adjustable
Monitored FIO2 is more than 0.07
below the set FiO2 for 30 seconds.
Accuracy: ±0.03
Flow Sensor
Error
Alarms-Non
Adjustable
Flow sensor cannot calibrate or the
internal wire is damaged or sensor is
disconnected.
Flow -
Inspiratory
Setting
Control Panel
Range: Ped/Infant: 1 to 100 L/min
(resolution 1 L/min)
Adult: 1 to 180 L/min (resolution 1 L/min)
Accuracy: ±10% or ± 0.3 L/min,
whichever is greater
Flow -
Inspiratory
Monitored
GUI-Monitored
Data
Displayed range: 0 to 200 L/min
(resolution 1 L/min)
Flow-
Expiratory
GUI Monitored
Data
Displayed range: 0 to 200 L/min
(resolution 1 L/min)
Flow
Waveform
GUI-Advanced
Data Set
Range: Square or Descending Ramp
Selects flow waveform for Volume
Control mandatory breaths.