Specifications
Item
Location and/or
Function
Range and Resolution or
Description
Freeze/Start/
Save
GUI-Main
and Extended
Functions
Freeze: Suspends plotting of graphs
(waveforms, loops, or trends) and
holds the current display for extended
viewing.
Start: Resumes plotting of graphs.
Save: Captures current screen image
for later downloading.
Gas Supply
Alarms-Non
Adjustable
If one or both gas supplies are below
30 psi.
High Baseline
Pressure
Alarms-Non
Adjustable
Monitored PEEP (Pbase) is greater
than set PEEP level by 5 cmH
2O/mbar
for two consecutive breaths.
Accuracy: ±1 cmH2O/mbar
High MV
E
GUI-Alarms-
Adjustable
Range:
Ped/Infant: 0.02 to 9.99 L (resolution
0.01 L)
10.0 to 60.0 L (resolution 0.1 L)
Adult: 2.00 to 9.99 L (resolution
0.01 L)
High Paw
GUI-Alarms-
Adjustable
Range:
Ped/Infant: 5 to 100 cmH
2O/mbar
(resolution 1 cmH2O/mbar)
Adult: 5 to 120 cmH2O/mbar
(resolution 1 cmH2O/mbar)
Accuracy: ±3% or ±2 cmH2O/mbar,
whichever is greater
NOTE: A High Paw alarm violation
terminates the current breath and
cycles to exhalation. This alarm is
applicable for all breaths, including
manual inflations.
High RRtot
GUI-Alarms-
Adjustable
Total respiratory rate including
mandatory and spontaneous breaths.
Range:
10-120 b/min or OFF
Accuracy: +3% or +2 b/min,
whichever is greater
I:E Inverse
Violation
Alarms-Non
Adjustable
Ventilator settings result in an I:E ratio
greater than 4:1
I:E Ratio
(Monitored)
GUI-Monitored
Data
Displayed range:
From 99:1 to 10:1 (resolution 1)
From 9.9:1 to 1:9.9 (resolution 0.1)
From 1:10 to 1:99 (resolution 1)