Specifications
8
Item
Location and/or
Function
Range and Resolution or
Description
Ideal Body
Weight
GUI-Patient
Setup
Range: 2-2202 lb (1-999 kg)
Inspiratory
Hold
GUI-Extended
Functions
Range: Up to 15 seconds
Generates an Inspiratory Hold
maneuver at the end of a mandatory
breath inspiration for as long as the
button is pushed.
Insp time too
long
Alarms-Non
Adjustable
Ventilator settings result in an
Inspiratory Time greater than 5
seconds
Insp time too
short
GUI-Alarms-Non
Adjustable
Ventilator settings result in an
Inspiratory Time less than:
Adult: 0.1 Seconds
Infant/Peds: 0.05 seconds
Internal
Battery
Indicators
Control Panel
Internal Battery LED on the Control
Panel lights and an audible signal
sounds every five minutes to indicate
that the ventilator is operating on
internal battery power. The Battery
Charge Level icon (located in the
top right area of the GUI) shows
the relative level of internal battery
power when ventilator is operating on
internal battery power.
Leak Comp
(Leak
Compensation)
GUI-Patient
Setup
Range: ON or OFF
ON = 3 – 8 L/min for Ped/Infant
patient selection
3 – 15 L/min for Adult patient selection
OFF = 3 L/min regardless of leaks/ no
leak
Non-Invasive Ventilation ON- Leak
Compensation is automatically
adjusted from 3-25 L/min.
Loops
GUI-Main
Loops combinations:
Flow/Volume, Volume/Pressure or
both on one screen