OPR360US H0711
8-8
8
Specifications
Item
Location and/or
Function
Range and Resolution or
Description
Low Baseline
Pressure
Alarms-Non
Adjustable
When the monitored proximal
pressure </= low baseline pressure
criteria for more than 0.5 seconds for
two consecutive breaths.
Accuracy: ±1 cmH
2O/mbar
Low Battery
Alarms-Non
Adjustable
Unsilenceable audible alarm sounds
when internal battery capacity has
dropped to 25% or less.
Low MV
E
GUI-Alarms-
Adjustable
Range:
Ped/Infant: 0.01 to 9.99 L (resolution
0.01 L) 10.0 to 30.0 L (resolution 0.1L)
Adult: 1.00 to 9.99 L (resolution
0.01 L) 10.0 to 50.0 L (resolution
0.1 L)
Accuracy: ±10% or ±0.1 L, whichever
is greater
NOTE: When Non Invasive is
activated this alarm may be set to
OFF.
NOTE: Low MVE alarm is suspended
for 60 seconds once the breathing
circuit is reconnected following
activation of Suction Disconnect.
Low Paw
GUI-Alarms-
Adjustable
Range:
Ped/Infant: 3 to 75 cmH
2O/mbar
(resolution 1 cmH2O/mbar)
Adult: 3 to 95 cmH2O/mbar (resolution
1 cmH2O/mbar)
Accuracy: ±3% or ±2 cmH2O/mbar,
whichever is greater
Low Paw
Below PEEP
GUI-
Informational
Message
The current Low Paw alarm setting is
lower than the PEEP setting.
Mains
Control Panel
Lights when the ventilator is supplied
with AC power.
Manual
Inflation
Control Panel
Range: Up to 5 seconds
Delivers a manual inspiration while the
button is held down.
Terminates when the button is
released or when a High Pressure
Alarm is violated, whichever comes
first.
MV
E
GUI-Monitored
Data
Displayed range:
(cont. next page)