OPR360US H0711
8-9
Specifications
8
Item
Location and/or
Function
Range and Resolution or
Description
MV
E (cont.)
Ped/Inf: 0.00 to 9.99 L (resolution
0.01 L)
Adult: 10.0 to 99.9 L (resolution 0.1 L)
Accuracy: ±10% or ±0.3 L, whichever
is greater
NOTE: The display will not be
updated if the exhalation flow sensor
is disconnected.
MV
I
GUI-Monitored
Data
Displayed range:
Ped/Inf: 0.00 to 9.99 L (resolution
0.01 L)
Adult: 10.0 to 99.9 L (resolution 0.1 L)
Accuracy: ±10% or ±0.3 L, whichever
is greater.
NIF
GUI-Technical
Generates NIF maneuver for
measuring airway pressure during
a maximum inspiratory effort. It
is available in all modes, with and
without Non Invasive Ventilation (NIV)
activated.
Non-Invasive
Control Panel
Range: ON (LED lights) or OFF
Available in all modes and breath
types. Leak Compensation is
automatically adjusted up to 25 L/min
when Non Invasive is ON.
No
Communications
Alarms-Non
Adjustable
No communications failure is a failure
of the SBC to send communication
to the Main Board or loss of monitor
processor.
Unsilenceable audible alarm.
Note: Ventilator continues to ventilate
at current control setting.
Numeric
GUI-Main
The Numeric screen displays all
monitored parameters on a single
screen, including Advance data
functions (except Volume Target).
Numeric table includes: Ppeak, Pplat,
Pmean, PEEP, Total PEEP, FIO2,
I:E, Peak Insp Flow, Peak Exp Flow,
WOBim, Cdyn effective, Cstat, RI, RE,
Time Const., RR tot, RR spont, RSBI,
t Insp, MVE spont, VTI, VTE, VTE %
Variance, MVI, MVE, Slope/Rise, Exp.
Threshold, Pause, Open Exh, Flow
Wave