OPR360U A0509
5-1
5
OPR360U A0509
5-1
5
OPR360U A0509
5-1
5
OPR360U A0509
5-1
5
OPR360US H0711
8-10
8
Specifications
Item
Location and/or
Function
Range and Resolution or
Description
O2 3 Min
Control Panel
When pressed delivers 100% oxygen
for 3 minutes. Pressing button again
turns off 100% oxygen delivery.
Indicator lights when O
2 3 min is in
effect.
O2 Sensor
GUI-Sensors
Can perform an oxygen sensor
calibration by delivering 100% oxygen
to the system. O2 Sensor calibration is
always performed when the O2 3 min
button is pressed. Can disable O2 sensor.
O2 Sensor
Disconnected
GUI-
Informational
Message
Notification message that the O2
sensor has been disconnected.
O2 Sensor
Error
Alarms-Non
Adjustable
Alarm message is displayed and
audible alarm sounds if the oxygen
sensor fails.
Open
Exhalation
Valve
GUI-Advanced
Data Set
Range: ON or OFF
Activates management of a partially
open exhalation valve. Coverts
Pressure Control mandatory breath
type to Biphasic Pressure Release
mandatory breath type (enables BPRV).
P0.1
GUI-Technical
Generates P0.1 measurement for
assessing the patient’s respiratory
drive. It is available in all modes
and breath types but not available
when Non Invasive Ventilation (NIV) is
activated.
Patient
Category
GUI-Patient
Setup
Range: Ped/Infant (pediatric/infant)
Adult (An icon located on the top left
corner of the GUI indicates category
selected.)
NOTE: Settable ranges for ventilator
parameters and alarms will vary
depending on the patient category
selected.
Patient
Trigger
Indicator
GUI-Monitored
Data
The area behind the patient category
icon at the top left side of the GUI
lights green to indicate that a patient
trigger has been activated.
Pause
(Inspiratory)
GUI-Advanced
Data Set
Range: Off, 0.1 – 2.0 seconds
(resolution 0.1 second)
Sets the duration of Pause at the
end of inspiration for volume control
mandatory breaths.