Specifications
8
Item
Location and/or
Function
Range and Resolution or
Description
PEEP/CPAP
(Baseline
Pressure-Set)
Control Panel
Range:
Ped/Infant: 0 to 30 cmH
2O/mbar
(resolution 1 cmH2O/mbar)
Adult: 0 to 45 cmH2O /mbar
(resolution 1 cmH2O/mbar)
Accuracy: ±10% or ±1 cmH2O/mbar,
whichever is greater.
Baseline
Pressure-
Monitored
GUI-Monitored
Data
Displayed range: 0 to 99.9 cmH2O/
mbar (resolution 0.1 cmH2O/mbar)
Accuracy: ±3% or ±2 cmH2O/mbar,
whichever is greater.
total PEEP
GUI-Monitored
Data
Displayed range: 0 to 99.9 cmH
2O/
mbar
Resolution: 0.1 cmH2O/mbar
Accuracy: Larger of ± 3% or 2
cmH2O/mbar
Maneuver–based with time stamp (for
up to 24 hours)
Total PEEP equals the sum of set
PEEP + AutoPEEP. Total PEEP is
updated immediately following an Exp
Hold.
Pmean
GUI-Monitored
Data
Displayed range: 0 to 140 cmH
2O/
mbar (resolution 0.1 cmH2O/mbar)
Accuracy: ±3% or ±2 cmH2O/mbar,
whichever is greater
The average pressure in the patient
breathing circuit for the past 30
seconds
Power
Shutdown
Alarms-Non
Adjustable
When ventilator is powered off an
audible alarm sounds. Silence by
pressing Alarm Silence button.
Ppeak
GUI-Monitored
Data
Displayed range: 0 to 140 cmH
2O/
mbar (resolution 0.1 cmH2O/mbar)
Accuracy: ±3% or ±2 cmH2O/mbar,
whichever is greater
Updated following each positive
pressure inflation
Pplat
GUI-Monitored
Data
Displayed range: 0 to 140 cmH
2O/
mbar (resolution 0.1 cmH2O/mbar)
Accuracy: ±3% or ±2 cmH2O/mbar,
whichever is greater
NOTE: Displays time stamped
numeric value (for up to 24 hours)
following a valid inspiratory hold
maneuver or following Pause that
results in a stable pressure level.