OPR360U A0509
5-1
5
OPR360U A0509
5-1
5
OPR360U A0509
5-1
5
OPR360U A0509
5-1
5
OPR360US H0711
8-12
8
Specifications
Item
Location and/or
Function
Range and Resolution or
Description
Pressure Bar
Graph
Control Panel
Displayed range: –10 to 120 cmH
2O/
mbar
Pressure Limit
Control Panel
Range:
Ped/Infant: 0 to 70 cmH2O/mbar
(resolution 1 cmH2O/mbar)
Adult: 0 to 80 cmH2O/mbar (resolution
1 cmH2O/mbar)
Accuracy: ±10% or ±1 cmH2O/mbar,
whichever is greater
Pressure Limit
Below PEEP
GUI-
Informational
Message
The current Pressure Limit setting is
lower than the PEEP/CPAP setting.
Pressure
Support
Control Panel
Range:
Ped/Infant: 0 to 50 cmH
2O/mbar
(resolution 1 cmH2O/mbar) above
PEEP/CPAP setting
Adult: 0 to 60 cmH2O/mbar (resolution
1 cmH2O/mbar) above PEEP/CPAP
setting
Accuracy: ±10% or +1 cmH2O/mbar,
whichever is greater
NOTE: Out of Range alarm occurs if
PEEP + PS > 60 cmH2O/mbar.
Pressure
Support +
PEEP > 60
GUI-
Informational
Message
Indicates that the operator is
attempting to set a Pressure Support
or PEEP level that is greater than the
sum of the two settings.
Pressure
Units
GUI-Technical/
Regional
Select cmH
2O (centimeters/water
pressure) or mbar (millibars).
RE
(cmH2O/mbar/
L/s) (Expiratory
resistance)
GUI-Monitored
Data
Displayed range:
0 to 999.9 cmH
2O/mbar/L/s
Accuracy: ±1 cmH2O/mbar /L/s
Resolution: 0.1 cmH2O/mbar/L/s
RE = exhalation time constant/ Cstat.
Maneuver–based with time stamp (for
up to 24 hours)
The exhalation time constant is the
slope of volume/flow loop during
exhalation. Cstat must be valid to
calculate RE. Calculated for time-
triggered breaths only.
Regional
GUI-Technical
• Altitude Range: 0 to 4,000 meters (0
to 13,124 feet)