OPR360US H0711
8-13
Specifications
8
Item
Location and/or
Function
Range and Resolution or
Description
Resp Rate
(Respiratory
Rate)
Control Panel
Range:
Ped/Infant: 1 to 120 b/min (resolution
1 b/min)
Adult: 1 to 80 b/min (resolution 1 b/
min)
Accuracy: ±1 b/min or ±10% of breath
period, whichever is less
RI
(cmH2O/mbar/
L/s)
(Inspiratory
resistance)
GUI-Monitored
Data
Displayed range: 0 to 999.9 cmH
2O/
mbar/L/s
Accuracy: ±1 cmH
2O/mbar/L/s
Maneuver–based with time stamp (for
up to 24 hours)
RI = (Ppeak – Pplat)/ end inspiratory
flow. Calculated for volume control
breaths only. Ppeak and Pplat must
be valid to calculate RI. RI is updated
only once immediately following
update of Pplat (Insp Hold) or Pause.
RR spont
(b/min)
(Spontaneous
respiratory
rate)
GUI-Monitored
Data
Displayed range: 0 to 999 b/min
Accuracy: Larger of ± 3% or ± 2 b/min.
RR tot
(b/min)
(Total
respiratory
rate)
GUI-Monitored
Data
Displayed range: 0 to 999 b/min
Accuracy: ± 3% or ± 2 b/min.,
whichever is greater.
RSBI
(b/min/L)
(Rapid Shallow
Breathing
Index)
GUI-Monitored
Data
Displayed range: 0 to 9999 b/min/L
Accuracy: ± 1 b/min/L
Spontaneous respiratory rate/exhaled
tidal volume ratio. (RSBI =
RRspont2 / MVEspont). RR spont and
MVE (spont) must be valid to calculate
RSBI.
Screen Files
GUI-Technical
Contains list of the last 200 saved
screen images (.bmp files). Files are
assigned an 8-digit file name that
includes a letter for file type (W for
Wave, L for loop, etc), the last 4 digits
of the serial number, and a 3 digit
sequential number.
Setting/Alarm
Limit out of
Range
Alarms- Non
Adjustable
Alarm or setting parameter out
of range for the selected patient
category.