Specifications
Item
Location and/or
Function
Range and Resolution or
Description
Sigh
GUI-Patient
Setup
Range: ON or OFF
In Volume Control A/CMV or SIMV
delivers 1 sigh breath every 100
breaths at tidal volume x 1.5.
During a sigh breath flow remains
at set value and inspiratory time is
lengthened.
Slope/Rise
GUI-Advanced
Data Set
Range: 1-19
Resolution: 1, where 1 is the slowest
pressurization and 19 is the fastest.
Slope/Rise sets the pressurization
gain for Pressure Control, Pressure
Support, Volume Targeted Pressure
Control, BPRV and Volume Targeted
Pressure Support breaths.
Suction
Disconnect
Alarm Feature
Holding down the Alarm Silence
button for one second or longer (until
the ventilator sounds a short beep)
enables the Suction Disconnect
Function.
Sustained
High Baseline
Pressure
Alarms- Non
Adjustable
Monitored circuit pressure Pbase has
been > 8 cmH
2O/mbar above set
PEEP/CPAP for over 6 seconds for
Ped/Infant or 10 seconds for Adult
patients. Machine pressure is also > 8
cmH2O /mbar above set PEEP/CPAP.
Accuracy: ±1 cmH2O /mbar
WARNING! Ventilation and triggering
are suspended and the exhalation
valve and emergency intake valve
open to vent pressure.
t Insp
(inspiratory
time)
Control Panel
GUI Monitored
Data
Range:
Ped/Infant: 0.10 to 3.00 seconds
(resolution 0.01 seconds)
Adult: 0.10 to 5.00 seconds (resolution
0.01 seconds)
Accuracy: ±0.05 seconds
Displayed range: 0 to 9.99 seconds
(resolution 0.01 seconds)
Accuracy: ±0.05 second
Updated following each spontaneous
or mandatory breath.
Tidal Volume *
Control Panel
Range:
Ped/Infant: 20 to 1000 mL (resolution
2 mL)
(cont. next page)