OPR360US H0711
8-15
Specifications
8
Item
Location and/or
Function
Range and Resolution or
Description
Tidal Volume *
(cont.)
Control Panel
Adult: 100 to 995 mL (resolution 5
mL), 1.00 to 3.00 L (resolution 0.01 L)
Accuracy: ±10% or ± 2 mL, whichever
is greater
* Gas delivery is Body Temperature
Pressure Saturation compensated:
Body temperature 37 ºC, steam
saturated gas, and ambient pressure.
Time
Constant
GUI Monitored
Data
Displayed range: 0 to 9.99
Accuracy: +/- 0.01 s
Represents the exhalation Time
Constant: slope of flow-volume
loop during exhalation. Reliable
measurements require adequate time
for complete exhalation. Displayed as
time in seconds.
Trends
GUI-Main
Accuracy: Larger of the individual
parameter accuracy, or 2% of the
selected full scale.
Trigger
[Press Trig
button above
the display to
select Flow or
P (pressure)]
Control Panel
P (pressure trigger sensitivity):
Range: 0.0 to -5.0 cmH
2O/mbar
(resolution 0.1 cmH2O/mbar)
Accuracy: ±10%.
Flow (trigger sensitivity)
Range:
Ped/Infant: 0.1 to 2.0 L/min (resolution
0.1 L/min)
Adult: 0.6 to 2.0 L/ min (resolution 0.1
L/min)
Accuracy: ±10% or ±0.1 L/min,
whichever is greater
NOTE: Automatically compensated
for bias flow/leak compensation.
Ventilation
Suspended
Informational
Message
Displayed when the Suction
Disconnect feature has been activated
and the patient is disconnected
from the ventilator (See Suction
Disconnect).
Volume Target
GUI-Advanced
Data Set
Range: ON or OFF
When enabled, turns on Volume
Target Pressure Control mandatory
(cont. next page)