Specifications
Item
Location and/or
Function
Range and Resolution or
Description
Volume Target
(cont.)
GUI-Advanced
Data Set
breath type for all mandatory breaths
and Volume Target Pressure Support
for all spontaneous breaths.
Minimum pressure control/support
is PEEP + 5 cmH
2O/mbar, maximum
pressure control/support is Pressure
Limit setting.
When this feature is ON, Open Exh
(BPRV) is not active.
Volume Target
Not Met
Alarms-Non
Adjustable
In Volume Target Pressure Control
breath type the set tidal volume
cannot be delivered within the set
Pressure Limit/Inspiratory Time.
Volume Units
GUI-Patient
Setup
Range:
mL or mL/kg or mL/lb
VT
E (mL)
(expiratory tidal
volume)
GUI-Monitored
Data
Displayed range: 0 to 3000 mL
(resolution 1 mL)
Accuracy: For set tidal volume
> 0.10 L , ±10% or ±0.02 L, whichever
is greater.
For set tidal volume < 0.10 L, ±20%
or +/- 0.002L, whichever is greater.
VTI (mL)
(Inspiratory
tidal volume)
GUI-Monitored
Data
Displayed range: 0 to 3000 mL
(resolution 1 mL)
Accuracy: For set tidal volume > 0.10
L , ±10% or ±0.02 L, whichever is
greater.
For set tidal volume < 0.10 L, ±20%
or +/- 0.002L, whichever is greater.
VT
E % Var
GUI-Monitored
Data
Displayed Range: 0 to 100%
(resolution: 1%)
Accuracy: +/- 10%
Percent difference between inspiratory
and expiratory tidal volumes.
Waveforms
GUI-Main
Waveform combinations:
Pressure / Time, Flow / Time, Volume
/ Time or combination of two can be
displayed on one screen
Weight Units
GUI-Patient
Setup And
Quick Setup
Range: lb or kg
WOBimp
(J/min)
GUI-Monitored
Data
Displayed range: 0 to 99.99 J/min