OPR360US H0711
5-8
5
Alarms
Alarm or
Device Alert
Message
• Priority
• Type
•
Lamp Color
Description
and/or
Criteria
Remedy
Back Up
Ventilation
• Medium
• Patient
• Yellow
Monitored MVE
</= Low MVE
alarm limit.
Check patient.
Evaluate settings/readjust
as necessary. Check
trigger sensitivity.
Check that the breathing
circuit is intact and
securely connected.
The alarm is recovered
when MVE rises to 10%
above the Low MVE alarm
limit.
NOTE: Back up ventilation
is suspended for 60
seconds after changing
any ventilator settings and
following circuit reconnect
after the Suction
Disconnect feature is
used.
Both Air/O2
Supply Loss
• High
• Ventilator
• Red
Inlet pressure
for both air and
oxygen supplies
has dropped
below minimum
supply pressure
required.
Restore both gas supplies
and/or provide alternate
ventilation immediately.
Check that air and oxygen
supplies are connected
and providing >/= 30 psig
at ventilator inlet.
The ventilator opens the
emergency intake and
exhalation valves to allow
the patient to breathe
room air.
Check Flow
Sensor
Board
• Medium
• Ventilator
• Yellow
The flow
sensor cable
on the flow
sensor board is
disconnected
from the main
board. The
ventilator can
only detect the
flow sensor
cable during
ventilation.
The flow sensor cable
must be re-connected
to the Main Board by
a qualified service
representative.
NOTE: Message cannot
be cleared with the Reset
button.