What to do if Olympus Scanner has insufficient probe contact?
- JjrichardsAug 4, 2025
If you're experiencing insufficient probe contact with your Olympus Scanner, it might be due to an improperly set probe holder. With the scanner removed from the pipe, try gently forcing the probe holder inward by hand to increase the preload of the wedge on the inspection surface. When reinstalling the scanner onto the pipe, make sure that the probe holder is not opened more than necessary.