66
Table 16 Scanner setup chart for larger pipes
Pipe outside diameter
Tail
setting
Probe-
holder
position
Number of links
in. mm Half Male Female Medium Long
2.50 to 2.60 63.5 to 66.0 1.5
Extended 1 22.60 to 2.70 66.0 to 68.6 2.5
2.70 to 2.80 68.6 to 71.1 4
2.80 to 2.86 71.1 to 72.6 1.5
Extended 1 1 22.86 to 2.95 72.6 to 74.9 2.5
2.95 to 3.12 74.9 to 79.2 4
3.12 to 3.19 79.2 to 81.0 1.5
Extended 2 1 2
3.19 to 3.28 81.0 to 83.3 2.5
3.28 to 3.39 83.3 to 86.1 4
3.39 to 3.45 86.1 to 87.6 1.5
Extended 1 2 1 23.45 to 3.55 87.6 to 90.2 2.5
3.55 to 3.67 90.2 to 93.2 4
3.67 to 3.73 93.2 to 94.7 1.5
Extended 2 3
3.73 to 3.82 94.7 to 97.0 2.5
3.82 to 3.94 97.0 to 101.3 4
3.94 to 3.99 100.1 to 101.3 1.5
Extended 1 1 1 13.99 to 4.08 101.3 to 103.6 2.5
4.08 to 4.19 103.6 to 106.4 4
4.19 to 4.25 106.4 to 108.0 1.5
Extended 2 1
4.25 to 4.33 108.0 to 110.0 2.5
4.33 to 4.43 110.0 to 112.5 4
4.43 to 4.49 112.5 to 114.0 1.5
Extended 1 2 14.49 to 4.58 114.0 to 116.3 2.5
4.58 to 4.67 116.3 to 118.6 4