EasyManua.ls Logo

Olympus OER-Elite - 5.2 Flow of inspection

Olympus OER-Elite
138 pages
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
128
5.2 Flow of inspection
OER-Elite OPERATION MANUAL
Ch.5
WARNING
During inspection, always wear appropriate personal protective equipment, such as
eyewear, face mask, moisture-resistant clothing, and chemical-resistant gloves that
fit properly and are long enough so that your skin and eyes is not exposed. All
personal protective equipment should be inspected before use and replaced
periodically before it is damaged.
On the first day you use the reprocessor, inspect it by following the procedure below.
5.2 Flow of inspection
1
Check that the reprocessor can be turned ON.
Section 5.3 on page 129
2
Check that no fluid leaks from the water supply line and inside
the reprocessor.
Section 5.4 on page 132
3
Check that the lid and its packing are free of irregularity.
Section 5.5 on page 133
4
Check that the connectors of reprocessor are free of irregularity.
Section 5.6 on page 134
5
Check that the connecting tubes and leak test air tube are free of
irregularity.
Section 5.7 on page 137
6
Check the remaining amount of detergent.
Section 5.8 on page 138
7
Check the remaining amount of alcohol.
Section 5.9 on page 139

Table of Contents