4.2 Auto leak test setting
79
OER-Elite OPERATION MANUAL
Ch.4
This setting is to set the timing of the auto leak test when this function is included in the reprocessing
process.
For the auto leak test in the reprocessing process, refer to “ Auto leak test” on page 221.
CAUTION
Always perform leak test in a sink before manual cleaning. If the endoscope has a
water leaking point, water may penetrate the scope during manual cleaning, in
which case the scope may break down.
4.2 Auto leak test setting
1 Select the “Auto Leak Test Setting” on the settings menu.