104
4.16 Auxiliary water tube cleaning setting
OER-Elite OPERATION MANUAL
Ch.4
When this setting is activated, an auxiliary water tube (MAJ-855) can be reprocessed simultaneously
with an endoscope with water jet function by using dedicated connecting tube MAJ-2138. In this case,
two endoscopes with two auxiliary water tubes can be reprocessed in one reprocessing process.
To obtain optional connecting tube MAJ-2138, contact Olympus.
WARNING
If the setting is activated, always use connecting tube MAJ-2138 for an endoscope
with water jet function. Otherwise, channel connectivity monitoring may not be able
to detect improper connection of the connecting tubes.
NOTE
If the setting is activated, the Connection guide screen which appears after reading
the scope ID displays MAJ-2138 as the connecting tube to be connected to the
water jet function.
If the setting is activated, an endoscope with water jet function cannot be
reprocessed without an auxiliary water tube (MAJ-855).
If this setting is disabled, either 2 endoscopes or 1 endoscope and 1 auxiliary water
tube (MAJ-855) can be simultaneously reprocessed.
For details how to install endoscopes and auxiliary water tubes simultaneously,
refer to Section 6.6, “Loading of endoscopes and accessories”.
4.16 Auxiliary water tube cleaning setting
1 Select “Auxiliary Water Tube Cleaning Setting” on the Setting menu.
2 Select setting from options. Then press “OK” or “Save” button.