The selected settings are displayed with white buttons.
2 Select setting from options.
No. Item/Button Setting options Description
1 ALT setting Beginning of the
process
Auto leak test is performed at the starting of “Reprocessing”. The
reprocessing process time is extended by about 2 minutes. If
“Beginning of the process” is selected and hot water is used for
manual cleaning, the ALT result may be “Leaked” even though the
endoscope has no leakage. The leak test result may become
“Leaked” depending on the temperature of the endoscope. To
perform auto leak test with the ultrasonic endoscope, wait for more
than 30 minutes. To perform it with the other type of endoscope,
wait for more than 15 minutes.
End of the process
(default)
Auto leak test is performed at the end of “Reprocessing”. The
reprocessing process time is extended by about 2 – 7 minutes.
When “End of the process” was chosen and there is a leak with an
endoscope, water may enter the endoscope from the leakage
point.
2 Cancel – Returns to the Setting menu without saving the selected setting.
3 Save – Returns to the Setting menu after saving the selected setting.
3 Press the “Save” button to save the selected setting.