Important Information — Please Read Before Use
15
OER-Elite OPERATION MANUAL
WARNING
Do not install this reprocessor in any place where any of the following are present.
High oxygen concentration
Oxidizing substance such as Nitrous Oxide (N
2
O)
Flammable anesthetic gas
This reprocessor is not explosion-proof and may explode or cause a fire under
these conditions.
Always use the power cord provided with this reprocessor. Otherwise, reprocessor
failure or power cord burnout may result. Also, remember that the provided power
cord is for use only with this reprocessor and should not be used with any other
equipment.
CAUTION
Do not press any of the switches on the control panel of this reprocessor with a
pointed or hard object. Otherwise, the switch may be damaged.
Be sure to turn off the water faucet and the power switch of the reprocessor at the
end of the day to avoid potential water leaks.
To avoid malfunctions, do not use this reprocessor in a dusty environment.
To avoid electromagnetic interference from other equipment, do not install any
other electrical devices in close proximity to this reprocessor (aside from ancillary
devices used with this reprocessor).
This equipment enables radio communication by RFID and emits RF (radio
frequency: 13.56 MHz) energy to perform the said intended functions. It may cause
electromagnetic interference in nearby electronic equipment, and is labeled with
the symbol below. If electromagnetic interference occurs, mitigation measures may
be necessary, such as moving the electronic equipment away, reorienting or
relocating this instrument, or shielding the location. An electromagnetic
interference with other devices may shorten the communications distance of the
designated ID tag and cause signals to become unreadable. Try to take mitigation
measures such as keeping the affecting device away from this equipment.