108
Chapter 6 Reprocessing, Storage, and Disposal
HIGH FLOW INSUFFLATION UNIT UHI-4
6.3 Compatible Reprocessing Methods
Compatibility summary
High Flow Insufflation Unit and the accessories are compatible with several
methods of reprocessing. For information concerning the applicable
reprocessing methods and parameters, refer to “List of compatible methods” on
page 109. Additionally, some reprocessing methods may cause degradation of
medical devices that shortens product life. For information concerning
degradation of medical devices from reprocessing and its number of times, refer
to Section 6.10, “Signs of degradation from reprocessing and its number of
times”. Follow the policies at your local institution when choosing which methods
to employ.
Methods listed as “compatible” in Table 6.1 are compatible
for routine use only when used according to the
manufacturer’s instructions. Repeated use and reprocessing
of High Flow Insufflation Unit and the accessories lead to
gradual wear and tear. Furthermore, reprocessing methods
that employ higher temperatures and more caustic/corrosive
materials may lead to faster deterioration. In general,
sterilization processes are harsher on equipment than
disinfection processes. Before each patient procedure,
inspect High Flow Insufflation Unit and the accessories for
damage, according to the instructions described in this
manual and its companion “OPERATION MANUAL”.
The Instructions provided in this manual regarding
compatible reprocessing methods are not valid for Olympus
devices repaired by a non-Olympus facility. Olympus repairs
devices to manufacturer’s specifications using original
equipment manufacturer’s (OEM) materials. The use of non-
OEM materials to repair an Olympus device may affect the
material compatibility and reprocessing efficacy of the device
with certain reprocessing chemicals or methods. In the event
that your device has been repaired by a non-Olympus facility,
contact the repair facility for instructions regarding compatible
reprocessing methods.