Chapter 6 Reprocessing, Storage, and Disposal
111
HIGH FLOW INSUFFLATION UNIT UHI-4
6.5 Surface disinfectant cleaner
Use a medical-grade hydrogen peroxide surface disinfectant cleaner with
properties as shown in Table 6.3.
Use the surface disinfectant cleaner cleared/approved by
your national regulatory agency.
Follow the surface disinfectant cleaner manufacturer’s instructions regarding
concentration, drying, temperature, contact time, use life, and expiration date.
The surface disinfectant cleaner shown in Table 6.4 was used for validation.
6.6 Water
General using for the reprocessing
Use either fresh, potable water or water that has been processed (e.g., filtered,
deionized, or purified) to improve its chemical and/or microbiological quality.
Consult with your institution’s infection control committee.
Surface disinfectant cleaner Hydrogen peroxide
Percentage solution Undiluted solution
Disinfectant concentration Hydrogen peroxide 1.5 g/100 g
Component Hydrogen peroxide, Glycolic acid
Others Acting as disinfectant and cleaner No rinsing required
Table 6.3 Hydrogen peroxide surface disinfectant cleaner with properties
Trade name Type Manufacture
Incidin
OxyFoam S
Hydrogen peroxide ECOLAB
Table 6.4 Surface disinfectant cleaner used for validation