80
Chapter 5 Operation
HIGH FLOW INSUFFLATION UNIT UHI-4
5.4 Setting the cavity pressure
Maintaining the cavity pressure at an excessively high
pressure for a long period may cause complications, etc.
Take sufficient care on the cavity pressure setting and always
use the optimum pressure.
Set the cavity pressure to the desired value using the “+” or “–” pressure setting
switch. Pressing “+” increases the set value and pressing “–” decreases it.
The available cavity pressure range varies depending on the
cavity mode.
To increase the pressure to 20 mmHg or 21 mmHg in the
“NORMAL” cavity mode, press and hold the “+” pressure
selector switch for 3 seconds.
Cavity mode Setting pressure range Note
NORMAL
3 – 25 mmHg
(0.4 – 3.3 kPa)
To increase the pressure from 20
to 21 mmHg, keep pressing “+”
for at least 3 seconds.
SMALL
3 – 15 mmHg
(0.4 – 2.0 kPa)
Table 5.2