44
Chapter 3 Installation and Connection
HIGH FLOW INSUFFLATION UNIT UHI-4
For the other market
1. Inspect the medical gas pipeline adapter for the high flow insufflation unit for
damage, cracks, and irregularities.
2. Use the supplied wrench to attach the medical gas pipeline adapter for the
high flow insufflation unit to the CO
2
gas inlet on the rear panel of the high
flow insufflation unit and securely tighten it. Tighten with a force of about
24.5 N·m (2.5 kgf/m) (see Figure 3.3 on page 41).
3. Connect the medical gas pipeline adapter to the medical gas pipeline hose
and tighten the adapter manually with a force (about 5 N·m (0.5 kgf/m)) until
the position where the connector is stopped (see Figure 3.7).
Figure 3.7
4. Connect the hose to the CO
2
connector of the medical gas pipeline.
3.5 Connecting the foot switch (MAJ-1939)
• Never connect any other equipment to the foot switch
connector.
• The connector section of the foot switch is not watertight.
Keep it away from fluids.
Medical gas pipeline hose
Medical gas pipeline adapter
for CO
2
regulation unit
Connector