Important Information — Please Read Before Use
HIGH FLOW INSUFFLATION UNIT UHI-4
Dangers, warnings, and cautions
Follow the dangers, warnings, and cautions given below when handling this
device.
This information is to be supplemented by the dangers, warnings, and cautions
given in each chapter.
• The UHI-4 is not explosion-proof. Never install or operate it in
the presence of flammable gases.
• Supply medical grade CO
2
gas only. Never use other kinds of
gas. Using gases other than CO
2
gas may result in fire,
poisoning, complications, etc. Use of non-medical grade, oil
polluted CO
2
might result in a malfunction of the insufflation
pressure control stage, causing a serious patient risk. Use
the hose for cylinder or adapter for medical gas pipeline as
described in this instruction manual to connect the CO
2
gas
cylinder or medical gas pipeline.
• Use of this device for cavity distention is contraindicated
whenever endoscopy is contraindicated. Refer to the
operator’s manual of your endoscope for absolute and
relative the contraindications.
• This device is contraindicated for hysteroscopic insufflation,
i.e., it must not be used for intrauterine distention.
• Strictly observe the following precautions. Failure to do so
may place the patient and medical personnel in danger of an
electric shock.
When the high flow insufflation unit is used to examine a
patient, do not allow metal parts of the endoscope or its
accessories to touch metal parts of other system
components. Such contact may cause unintended current
flow to the patient.
Keep fluids away from all electrical equipment. If fluids are
spilled on or into the high flow insufflation unit, stop
operation of the unit immediately and contact Olympus.
Do not prepare, inspect, or use this high flow insufflation
unit with wet hands.