118
Chapter 6 Reprocessing, Storage, and Disposal
HIGH FLOW INSUFFLATION UNIT UHI-4
High Flow Insufflation Unit (UHI-4)
Reprocessing may cause the following degradations. If any of these
signs of degradation happens, contact Olympus.
Crack appears on the front panel or the Power switch.
Discoloration on the front panel or the top cover.
Peeling of the labels on the front panel.
It has been confirmed that there are no degradation events
due to reprocessing during the service life.
Smoke Evacuation Foot Switch (MAJ-1939)
It has been confirmed that there are no degradation events due to
reprocessing during the service life.
Accessory (MAJ-590, MAJ-591)
Reprocessing may cause the cracks or discoloration. If any of these signs of
degradation happens, discard and replace these accessories.
Olympus has verified no degradation happens when the Table 6.8 listed
number of reprocessing cycle is performed. When this accessory is
reprocessed more times than the following times, cracks and unacceptable
discoloration may occur.
Insufllation Tube (MAJ-590) Suction Tube (MAJ-591)
Detergent 50 times 50 times
Washer Disinfector 50 times 50 times
Steam sterilization
(autoclaving)
50 times 50 times
Table 6.8